The Edwards E2M18 / A36315903 vacuum pumps are direct drive, sliding vane pumps. The E2M18 is a dual-stage vaccuum pump.
The Edwards E2M18 pump is oil-sealed and designed for reliable, long-term operation in both laboratory and industrial environments. The pump is a free-standing unit. The drive is provided through a flexible coupling by a single-phase or three-phase (four pole) motor. The motors are totally enclosed and are cooled by the motor-cooling fan which directs air along the motor fins.
This pump has been rebuilt with Hydrocarbon oil.
The Edwards single-phase motors have a thermal overload device. When the motor is too hot, the thermal overload device switches off the pump. The thermal overload device has an automatic reset; when the motor cools down, the device resets and the motor will restart. The single phase motors have an on/off switch.
An oil pressure system lubricates the pump shaft bearing surfaces and rotor sealing faces. The pump has an oil distribution valve which prevents discharge of oil to the pump interior (suckback) after the pump stops. The pumping chambers are air-tight, so this arrangement prevents air suckback unless the gasballast valve is open.
You can inspect the level and condition of oil in the oil box reservoir through the oil sight-glass. An oil filler-plug is fitted to the top of the oil box. The pump has an oil drain-plug to allow gravity oil drain. It also has an inlet-port, outlet nozzle and gas-ballast control. The pump is mounted on an aluminium alloy baseplate on rubber pads.
When you use the pump with high vapour throughputs, you should use the gas-ballast facility to prevent condensation of the vapours inside the pump. The condensates will contaminate the oil, will cause performance to deteriorate and may cause corrosion of the pump mechanism.
A&J Vacuum also repair / rebuild and service all Edwards E1M18, E2M18, E2M18, E2M28 Series Vacuum Pumps along with all other Edwards Turbo Pumps, Edwards Dry Pumps, Edwards Scroll Pumps and all other Edwards Vacuum Rotary Vane pumps. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
Power Total Rated @60Hz | 950.00 W |
Power Total Rated @50Hz | 750.00 W |
Pumping Speed Air Nitrogen 60Hz | 12.01 cfm |
Pumping Speed Air Nitrogen 50Hz | 10.01 cfm |
Pressure Water Vapour Max | 20.00 mbar |
Nominal Rotational Speed 50Hz | 24.00 Hz |
Capacity Water Vapour | 8.300 × 10⁻⁵ kg/s |
Displacement @50Hz Max | 12.07 cfm |
Pressure Ultimate With Gas Ballast | 0.015 mbar |
Pressure Ultimate | 1.000 × 10⁻³ mbar |
Nominal Rotational Speed 60Hz | 28.67 Hz |
Displacement @60Hz Max | 14.71 cfm |
Power at Ultimate | 750.00 W |
Connection Inlet Flange | DN 25 ISO-KF (NW25) |
Weight | 39.00kg / 85 lbs |
Connection vacuum outlet flange | 15 mm nozzle | 3/4" BSPP (or G 3/4") Female |
Power Supply Phase |
1-Ph |
Power Supply Voltage | 1Ph 220-240V 50/60Hz |
Dimension Length | 22.76 in |
IP Rating | IP44 |
Power Supply Frequency | 50/60HZ |
Dimension Height | 10.67 in |
Dimension Width | 7.09 in |
Connector Electrical Power | Hard wiring (in electrical box of pump / mains) |
Noise / Sound Level | 57.00 db(A) |
Cooling Method | Air Cooling (forced - fan) |
Motor Efficiency Class | IE2 |
Connection Inlet/Outlet Flange | DN 25 ISO-KF (NW25) / 15 mm nozzle |
Certification Motor | UL | CSA |
Temperature Ambient Operation Max | 104.00°F | 40.00°C | 313.15 K |
Temperature Ambient Operation Min | 55.40°F | 13.00°C | 286.15 K |
Temperature Ambient Storage Min | -22.00°F | -30.00°C | 243.15 K |
Temperature Ambient Storage Max | 158.00°F | 70.00°C | 343.15 K |
Classification Product Pressure Range | Medium Vacuum: 1 mbar - 1·10⁻³ mbar Rough Vacuum: 1000 mbar - 1 mbar |
Classification High / Medium / Rough Vacuum pump | Medium Vacuum Pump Rough Vacuum Pump |
Classification Dry or Wet Pump | Wet |
Classification Vacuum Pump Technology | Classification Vacuum Pump Technology |
Default Oil Type | ULTRAGRADE™ Performance 19 |
Altitude Max | 78740.16 in |
Oil Capacity Min | 7.500 × 10⁻⁴ m³ |
Oil Capacity Max | 1.050 × 10⁻³ m³ |
Pressure Absolute Outlet Continous Max | 1500.00 mbar |
Material Exposed to Vacuum | High Grade Cast Iron, cast aluminium, fluorocarbon elastomer, nitrile, silicon, chemically resistant polymers, nickel , stainless steel |
Material Rotor | High Grade Cast Iron |
Material Stator | High Grade Cast Iron |
A&J Vacuum also repair / rebuild and service all Edwards E1M18, E2M18, E2M18, E2M28 Series Vacuum Pumps along with all other Edwards Turbo Pumps, Edwards Dry Pumps, Edwards Scroll Pumps and all other Edwards Vacuum Rotary Vane pumps. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.
The Edwards E2M18 / A36315903 vacuum pumps are direct drive, sliding vane pumps. The E2M18 is a dual-stage vaccuum pump.
The Edwards E2M18 pump is oil-sealed and designed for reliable, long-term operation in both laboratory and industrial environments. The pump is a free-standing unit. The drive is provided through a flexible coupling by a single-phase or three-phase (four pole) motor. The motors are totally enclosed and are cooled by the motor-cooling fan which directs air along the motor fins.
This pump has been rebuilt with Hydrocarbon oil.
The Edwards single-phase motors have a thermal overload device. When the motor is too hot, the thermal overload device switches off the pump. The thermal overload device has an automatic reset; when the motor cools down, the device resets and the motor will restart. The single phase motors have an on/off switch.
An oil pressure system lubricates the pump shaft bearing surfaces and rotor sealing faces. The pump has an oil distribution valve which prevents discharge of oil to the pump interior (suckback) after the pump stops. The pumping chambers are air-tight, so this arrangement prevents air suckback unless the gasballast valve is open.
You can inspect the level and condition of oil in the oil box reservoir through the oil sight-glass. An oil filler-plug is fitted to the top of the oil box. The pump has an oil drain-plug to allow gravity oil drain. It also has an inlet-port, outlet nozzle and gas-ballast control. The pump is mounted on an aluminium alloy baseplate on rubber pads.
When you use the pump with high vapour throughputs, you should use the gas-ballast facility to prevent condensation of the vapours inside the pump. The condensates will contaminate the oil, will cause performance to deteriorate and may cause corrosion of the pump mechanism.
A&J Vacuum also repair / rebuild and service all Edwards E1M18, E2M18, E2M18, E2M28 Series Vacuum Pumps along with all other Edwards Turbo Pumps, Edwards Dry Pumps, Edwards Scroll Pumps and all other Edwards Vacuum Rotary Vane pumps. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
Power Total Rated @60Hz | 950.00 W |
Power Total Rated @50Hz | 750.00 W |
Pumping Speed Air Nitrogen 60Hz | 12.01 cfm |
Pumping Speed Air Nitrogen 50Hz | 10.01 cfm |
Pressure Water Vapour Max | 20.00 mbar |
Nominal Rotational Speed 50Hz | 24.00 Hz |
Capacity Water Vapour | 8.300 × 10⁻⁵ kg/s |
Displacement @50Hz Max | 12.07 cfm |
Pressure Ultimate With Gas Ballast | 0.015 mbar |
Pressure Ultimate | 1.000 × 10⁻³ mbar |
Nominal Rotational Speed 60Hz | 28.67 Hz |
Displacement @60Hz Max | 14.71 cfm |
Power at Ultimate | 750.00 W |
Connection Inlet Flange | DN 25 ISO-KF (NW25) |
Weight | 39.00kg / 85 lbs |
Connection vacuum outlet flange | 15 mm nozzle | 3/4" BSPP (or G 3/4") Female |
Power Supply Phase |
1-Ph |
Power Supply Voltage | 1Ph 220-240V 50/60Hz |
Dimension Length | 22.76 in |
IP Rating | IP44 |
Power Supply Frequency | 50/60HZ |
Dimension Height | 10.67 in |
Dimension Width | 7.09 in |
Connector Electrical Power | Hard wiring (in electrical box of pump / mains) |
Noise / Sound Level | 57.00 db(A) |
Cooling Method | Air Cooling (forced - fan) |
Motor Efficiency Class | IE2 |
Connection Inlet/Outlet Flange | DN 25 ISO-KF (NW25) / 15 mm nozzle |
Certification Motor | UL | CSA |
Temperature Ambient Operation Max | 104.00°F | 40.00°C | 313.15 K |
Temperature Ambient Operation Min | 55.40°F | 13.00°C | 286.15 K |
Temperature Ambient Storage Min | -22.00°F | -30.00°C | 243.15 K |
Temperature Ambient Storage Max | 158.00°F | 70.00°C | 343.15 K |
Classification Product Pressure Range | Medium Vacuum: 1 mbar - 1·10⁻³ mbar Rough Vacuum: 1000 mbar - 1 mbar |
Classification High / Medium / Rough Vacuum pump | Medium Vacuum Pump Rough Vacuum Pump |
Classification Dry or Wet Pump | Wet |
Classification Vacuum Pump Technology | Classification Vacuum Pump Technology |
Default Oil Type | ULTRAGRADE™ Performance 19 |
Altitude Max | 78740.16 in |
Oil Capacity Min | 7.500 × 10⁻⁴ m³ |
Oil Capacity Max | 1.050 × 10⁻³ m³ |
Pressure Absolute Outlet Continous Max | 1500.00 mbar |
Material Exposed to Vacuum | High Grade Cast Iron, cast aluminium, fluorocarbon elastomer, nitrile, silicon, chemically resistant polymers, nickel , stainless steel |
Material Rotor | High Grade Cast Iron |
Material Stator | High Grade Cast Iron |
A&J Vacuum also repair / rebuild and service all Edwards E1M18, E2M18, E2M18, E2M28 Series Vacuum Pumps along with all other Edwards Turbo Pumps, Edwards Dry Pumps, Edwards Scroll Pumps and all other Edwards Vacuum Rotary Vane pumps. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.